IS 10053:1981 specifies the requirements for equipment used in the Jackson candle turbidimeter method and the procedure for determining turbidity in water samples. It details the design and calibration of the standard candle, turbidity tube, and supporting apparatus, enabling measurement of turbidity between 25 to 1000 Jackson Turbidity Units (JTU). This standard is essential for water quality engineers and laboratories conducting visual turbidity assessments using the Jackson candle method.
Overview
IS 10053:1981 specifies the requirements for equipment used in the Jackson candle turbidimeter method and the procedure for determining turbidity in water samples. It details the design and calibration of the standard candle, turbidity tube, and supporting apparatus, enabling measurement of turbidity between 25 to 1000 Jackson Turbidity Units (JTU). This standard is essential for water quality engineers and laboratories conducting visual turbidity assessments using the Jackson candle method.
Audience
Contents
Structure
IS 10053: Scope - Key Specifications & Tables
Scope:
IS 10053 covers methods for turbidity estimation in water using visual methods like candle turbidimeter or bottle standards.
Turbidity Estimation Methods:
Units:
Turbidity is expressed in Turbidity Units (TU).
Rounding off results:
Final test values are rounded as per JS: 2-1960, maintaining significant figures consistent with specified values.
| Light Path (cm) | Turbidity Units (TU) | Light Path (cm) | Turbidity Units (TU) |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 5.5 | 400 | 21.5 | 100 |
| 7.3 | 300 | 39.8 | 50 |
| 8.7 | 250 | 72.9 | 25 |
Light path measured from inside bottom of glass tube.
| Quantity | Unit | Symbol |
|---|---|---|
| Length | metre | m |
| Mass | kilogram | kg |
| Time | second | s |
| Electric current | ampere | A |
| Thermodynamic temp. | kelvin | K |
| Luminous intensity | candela | cd |
| Amount of substance | mole | mol |
| Force | newton | N |
| Pressure, stress | pascal | Pa |
flowchart LR
A[Water Sample] --> B[Candle Turbidimeter]
B --> C[Light Path Measurement]
C --> D[Compare with Standard Turbidity Units Table]
D --> E[Estimate Turbidity (TU)]
This scope ensures consistent turbidity measurement with internationally coordinated units
IS 10053: Terminology & Key Specifications
| Light Path (cm) | Turbidity Units | Light Path (cm) | Turbidity Units |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 5.5 | 400 | 21.5 | 100 |
| 8.7 | 250 | 72.9 | 25 |
flowchart LR
A[Standard Candle] --> B[Candle Holder Cover]
B --> C[Annular Extension Tube]
C --> D[Glass Tube with Sample]
D --> E[Light Path Measurement]
E --> F[Turbidity Units Determined]
This concise overview helps understand the terminology, units, and turbidity measurement standards per IS
Jackson Turbidity Units (JTU) & Measurement Range (IS 10053)
Definition: Turbidity measured by the candle turbidimeter expressed in JTU (Clause 3.1).
Measurement Range:
Calibration Standard:
Recording Turbidity (Table B-3.1):
| Turbidity Range (JTU) | Recording Precision (Nearest JTU) |
|---|---|
| 0 - 1.0 | 0.1 |
| 1 - 10 | 1 |
| 10 - 100 | 5 |
| 100 - 400 | 10 |
| 400 - 700 | 50 |
| 700 or more | 100 |
flowchart LR
A[Sample Preparation] --> B[Shake Sample & Standards]
B --> C[Place in Identical Bottles]
C --> D[Visual Comparison Against Standards]
D --> E[Record Turbidity as Closest Standard]
This method ensures reliable turbidity readings within the specified JTU ranges.
IS 10053: Equipment Requirements for Jackson Candle Turbidimeter
Pressure (Pa):
[
1 \text{ Pa} = 1 \text{ N/m}^2 = 1 \text{ kg·m}^{-1}\text{s}^{-2}
]
Force (N):
[
1 \text{ N} = 1 \text{ kg·m/s}^2
]
| Quantity | Unit | Symbol | Definition |
|---|---|---|---|
| Length | metre | m | |
| Luminous intensity | candela | cd | |
| Force | newton | N | 1 N = 1 kg·m/s² |
| Pressure, stress | pascal | Pa | 1 Pa = 1 N/m² |
| Frequency | hertz | Hz | 1 Hz = 1 cycle/s |
flowchart LR
A[Jackson Candle] --> B[Turbidity Estimation]
C[Bottle Standards] --> B
B --> D[Visual Comparison]
D --> E[Result in JTU]
Note: For detailed calibration and operational procedures, refer to
Candle Dimensions & Material (Clause 4.2):
Candle Holder & Support:
| Light Path (cm) | Turbidity Units | Light Path (cm) | Turbidity Units |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 4.5 | 500 | 18.0 | 120 |
| 7.0 | 310 | 36.7 | 55 |
| 8.7 | 250 | 72.9 | 25 |
(Measured from inside bottom of glass tube)
flowchart TB
A[Standard Candle] --> B[Burn Rate 7.39-8.16 g/hr]
B --> C[Flame: Clear, White, Smokeless]
A --> D[Dimensions: 150 mm x 20 mm]
A --> E[Material: Paraffin/Beeswax + Spermaceti]
F[Spring Holder] --> A
G[Glass Tube] --> H
IS 10053: Turbidity Tube Dimensions & Calibration
| Light Path (cm) | Turbidity Units | Light Path (cm) | Turbidity Units |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 5.5 | 400 | 22.6 | 95 |
| 7.8 | 280 | 43.5 | 45 |
| 8.7 | 250 | 72.9 | 25 |
Light path is measured from inside bottom of the tube.
flowchart TD
A[Turbidity Tube] --> B[Transparent Glass, 25mm dia]
B --> C[Length: 250-760 mm]
C --> D[Graduated 0-76 cm in 1 mm]
D --> E[Light Path Measurement]
E --> F[Calibration Table]
F --> G[Turbidity Units]
This setup ensures reliable turbidity measurement per IS 10053.
IS 10053: Tripod Stand and Support Assembly for Candle Turbidimeter
| Light Path (cm) | Turbidity Units | Light Path (cm) | Turbidity Units |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 4.5 | 500 | 22.6 | 95 |
| 7.8 | 280 | 43.5 | 45 |
| 8.7 | 250 | 72.9 | 25 |
(Light path measured from inside bottom of glass tube)
graph TD
A[Cast Iron Base] --> B[Brass Legs]
B --> C[Brass Candle Holder (Chromium Plated)]
C --> D[Spring Loaded Cylinder Support]
D --> E[Standard Candle]
F[Turbidity Tube] -->|Centreline aligned| E
E -->|Vertical distance 76 mm| F
This assembly ensures precise vertical alignment and consistent measurement conditions for turbidity testing using the candle turbidimeter.
IS 10053: Preparation of Standard Suspensions for Visual Comparison
| Light Path (cm)* | Turbidity (NTU) | Light Path (cm)* | Turbidity (NTU) |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 2.6 | 900 | 12.0 | 180 |
| 2.9 | 800 | 12.7 | 170 |
| 3.2 | 700 | 13.5 | 160 |
| 3.5 | 650 | 14.4 | 150 |
| 4.1 | 550 | 15.4 | 140 |
| 4.9 | 450 | 18.0 | 120 |
| 5.5 | 400 | 21.5 | 100 |
| 6.1 | 360 | 25.1 | 85 |
| 7.3 | 300 | 31.8 | 65 |
| 8.7 | 250 | 43.5 | 45 |
| 10.0+ | ... | ... | ... |
*Light path measured from inside bottom of glass tube.
Components:
Dimensions: All in millimeters (refer to IS 10053
IS 10053: Measurement Procedure for Turbidity
| Light Path (cm)* | Turbidity Units | Light Path (cm)* | Turbidity Units |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 5.5 | 400 | 21.5 | 100 |
| 7.3 | 300 | 39.8 | 50 |
| 8.7 | 250 | 72.9 | 25 |
*Measured from inside bottom of glass tube.
This table correlates light path length in the turbidimeter with turbidity units (NTU) for visual comparison.
flowchart TD
A[Prepare Sample] --> B[Fill Turbidimeter Tube]
B --> C[Observe Candle Flame]
C --> D[Compare Visibility with Table 1]
D --> E[Record Turbidity in NTU]
This concise procedure ensures reliable turbidity estimation per IS 10053.
IS 10053: Dilution & Calculation for High Turbidity Samples
Standard Suspension Preparation:
Dilution for High Turbidity (>1000 JTU):
Example:
Standard Turbidity Suspension:
[ \text{Turbidity}{original} = \text{Turbidity}{diluted} \times \text{Dilution Factor} ]
Where:
[ \text{Dilution Factor} = \frac{\text{Volume of sample} + \text{Volume of diluent}}{\text{Volume of sample}} ]
| Parameter | Value/Description |
|---|---|
| Standard suspension turbidity | 400 NTU |
| Turbidity-free water turbidity | 40 NTU |
| Max measurable turbidity without dilution | 1000 JTU |
| Dilution example | 1 part sample + 5 parts water |
| Resulting original turbidity | 3000 JTU (from example) |
flowchart LR
A[Sample with turbidity > 1000 JTU] --> B[Dilute with turbidity-free water]
B --> C[Measure diluted turbidity < 1000 JTU]
C --> D[Calculate original turbidity]
D -->|Original turbidity = Diluted turbidity × Dilution factor|
IS 10053: Stock & Standard Turbidity Suspensions (Appendix A)
| Suspension Type | Preparation Volume | Dilution | Turbidity (NTU) |
|---|---|---|---|
| Stock Turbidity | Mix 5 ml each sol. I & II in 100 ml flask | Stand 24h, dilute to mark | 400 |
| Standard Turbidity | 10 ml stock diluted to 100 ml | Dilution with turbidity-free water | 40 |
flowchart TD
A[Solution I (5 ml)] --> C[Mix in 100 ml flask]
B[Solution II (5 ml)] --> C
C --> D[Stand 24h at 25±3°C]
D --> E[Dilute to 100 ml mark & mix]
E --> F[Stock Suspension: 400 NTU]
F --> G[Dilute 10 ml to 100 ml]
G --> H[Standard Suspension: 40 NTU]
This ensures consistent turbidity standards for water testing per IS 10053.
IS 10053: Preparation of Standard Suspensions for Visual Comparison
| Light Path (cm) | Turbidity Units (NTU) | Light Path (cm) | Turbidity Units (NTU) |
|---|---|---|---|
| 2.3 | 1000 | 11.4 | 190 |
| 5.5 | 400 | 21.5 | 100 |
| 7.3 | 300 | 43.5 | 45 |
| 8.7 | 250 | 72.9 | 25 |
flowchart LR
A[Prepare Stock Suspension] --> B[Dilute to 100 ml]
B --> C[Obtain 400 NTU Suspension]
C --> D[Use Table for Turbidity Units vs Light Path]
D --> E[Visual Comparison in Turbidimeter]
This ensures standardized visual turbidity measurement per IS 10053.
Frequently Asked
Material and Dimensional Requirements for Standard Candle (IS 10053, Clause 4.2):
Material:
Dimensions:
Additional Features:
| Parameter | Specification |
|---|---|
| Material | Paraffin/beeswax + spermaceti |
| Solidifying Point | 58–60°C |
| Length | 150 mm |
| Diameter | 20 mm |
| Burn Rate | 7.39–8.16 g/hr (114–126 grains/hr) |
| Spring Material | Spring steel (IS:4454 Part I) |
| Flame Characteristics | Clear, white, smokeless, odorless |
Loading diagram...
This ensures consistent light source quality for turbidity measurement per IS 10053.
Calibration and Size Specifications of Turbidity Tube (IS 10053):
Calibration:
Size Specifications:
This ensures accurate visual turbidity measurement using the standard candle turbidimeter setup.
Loading diagram...
Procedure to Prepare Standard Turbidity Suspensions (IS 10053):
Stock Suspension Preparation (400 NTU):
Standard Suspension (40 NTU):
Comparison Procedure (5 to 100 JTU range):
| Suspension Type | Volume Mixed/Diluted | Final Volume | Turbidity (NTU) |
|---|---|---|---|
| Stock Suspension | 5 ml Solution I + 5 ml Solution II | 100 ml | 400 |
| Standard Suspension | 10 ml Stock + 90 ml water | 100 ml | 40 |
Loading diagram...
This method ensures reproducible turbidity standards for visual comparison in water testing.
The Jackson candle turbidimeter measures turbidity visually by observing a standard candle flame through a water sample. According to IS 10053:
| Parameter | Details |
|---|---|
| Measurement Principle | Visual disappearance of candle flame through suspension |
| Unit | Jackson Turbidity Units (JTU) |
| Direct Measurement Range | 25 JTU to 1000 JTU |
| Above 1000 JTU | Requires dilution |
| Below 25 JTU | Use indirect methods |
Loading diagram...
Steps for Measuring Turbidity Above 1000 JTU Using Dilution (IS 10053, Clause 2.1.2):
Example:
| Parameter | Value |
|---|---|
| Volume of sample | 1 volume |
| Volume of turbidity-free water added | 5 volumes |
| Measured turbidity of diluted sample | 500 JTU |
| Original turbidity | 500 JTU × (1 + 5) = 3000 JTU |
| Turbidity Range (JTU) | Nearest Unit to Record |
|---|---|
| 0 - 1.0 | 0.1 |
| 1 - 10 | 1 |
| 10 - 100 | 5 |
| 100 - 400 | 10 |
| 400 - 700 | 50 |
| 700 or more | 100 |
Loading diagram...
This method ensures accurate turbidity measurement beyond the direct range of the candle turbidimeter.
Ask AI about any clause, requirement, or provision in IS 10053. Get instant, clause-cited responses powered by our indexed library.
Free tier includes 150 queries (50 AI + 100 Reference) · No credit card required