eyepiece and screen graticules for determination of particle size of powders
1969 Edition

IS 5257:1969 outlines the specifications for designing and manufacturing eyepiece and screen graticules used in optical microscopes to measure particle sizes of powders, especially those passing a 75-micron sieve. This standard is indispensable for laboratories and industries analyzing powder particle size, guaranteeing precise and consistent sizing through calibrated grids and reference circles.

10Sections
24Clauses Indexed
AI Search Ready
1969Edition
Sieves Sieving and other Sizing MethodsCategory
Alternative search terms: eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 pdf free download, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 free download pdf, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969:1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969-1969 PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 (1969) PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 1969 edition PDF, eyepiece-and-screen-graticules-for-powder-particle-size-measurement-1969 edition 1969 PDF

What This Standard Covers

IS 5257:1969 outlines the specifications for designing and manufacturing eyepiece and screen graticules used in optical microscopes to measure particle sizes of powders, especially those passing a 75-micron sieve. This standard is indispensable for laboratories and industries analyzing powder particle size, guaranteeing precise and consistent sizing through calibrated grids and reference circles.

Who Uses This Standard

  • Laboratory Technicians for Material Testing
  • Quality Assurance Engineers in Powder Manufacturing
  • Material Science Researchers
  • Microscopy Experts
  • Pharmaceutical Quality Analysts
  • Chemical Process Engineers
  • Metallurgical Specialists

Key Topics Covered

Specifications for eyepiece and screen graticule design
Layout and dimensions of rectangular grids and circular patterns
Calibration markings and allowable measurement deviations
Materials and marking protocols for graticule fabrication
Utilization of graticules in optical microscopy for particle sizing
Measurement units and dimensional precision
Adaptation for powders passing through 75-micron sieves
Quality control and standard compliance markings
Comparison with international benchmarks like BS 3625:1963
Implementation in particle size distribution evaluation
Permissible line widths and circle diameter tolerances
Guidelines for grid length and width parameters

Table of Contents

1Scope and Essential Specifications

IS 5257 - Scope and Fundamental Criteria for Graticules

Scope

  • Pertains to eyepiece and screen graticules employed in determining particle size of powders that pass through a 75-micron IS sieve.
  • Designed for use with optical microscopes where particle images are compared against calibrated reference circles.
  • Ensures conformity with standards such as BS 3625:1963 for consistency.

Principal Specifications (Clause 2.3.1 & Table 1)

ComponentValue & Allowable Deviation (units = 1/100th grid length)
Grid length100 (reference unit)
Grid breadth50 ± 0.25
Spacing between calibration marks85.4 ± 0.25
Circle Diameters:
- Circle 11.41 ± 0.03
- Circle 22.00 ± 0.04
- Circle 32.83 ± 0.06
- Circle 44.00 ± 0.08
- Circle 55.66 ± 0.10
- Circle 68.00 ± 0.12
- Circle 711.31 ± 0.14

Additional Details

  • Boundary lines extend 2.5 units to intersect.
  • Calibration marks on the longer axis are 7.5 units in length, symmetrically positioned.
  • Ruling line thickness is ≤ 1/5 unit (max 1/500th grid length).
  • Measurements between centers of ruling lines; circle diameters measured at outside edges.

flowchart LR
    A[Powder Sample] --> B[Microscope with Graticule]
    B --> C[Particle Image Compared to Reference Circles]
    C --> D[Particle Size Distribution Determination]

This standard guarantees accurate and reproducible particle size measurements using standardized graticules.

2Graticule Construction Details

IS 5257: Constructional Requirements for Eyepiece and Screen Graticules

Core Specifications (Clause 2.3.1 & Table 1):

  • Grid Dimensions:
    • Length: 100 units
    • Breadth: 50 ± 0.25 units
  • Boundary Lines: Extend by 2.5 units beyond grid edges to form crosses.
  • Calibration Marks: Positioned on the longer axis with a length of 7.5 units, symmetrically arranged.
  • Distance Between Calibration Marks: 85.4 ± 0.25 units
  • Ruling Line Thickness:1/5 unit (max 1/500th of grid length), consistent within ±20%.
  • Circle Diameters (units ± tolerance):
Circle No.Diameter (units)Tolerance
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14
  • Lengths are measured center to center of ruling lines; circle diameters between outer edges.

Purpose and Application

  • Graticules are utilized for determining particle sizes of powders passing through a 75-micron sieve.
  • Particles are visually compared under microscopes against reference circles to ascertain size.

Diagrammatic Overview

graph TD
    A[Grid Boundary Lines] -->|Extend 2.5 units| B[Crossing Lines]
    B --> C[Rectangular Grid 100 x 50 units]
    C --> D[Calibration Marks 7.5 units symmetrically positioned]
    C --> E[Subdivision Lines Creating Defined Areas]
    C --> F[Open Circles with Specified Diameters]

Reference: IS 5257 (1969) — Specification for eyepiece and screen graticules used in particle size measurement.

2.1General Design Criteria for Graticules

IS 5257: General Design Principles for Graticules

Principal Specifications (Clause 2.3.1 & Table 1)

  • Grid Dimensions:
    • Length: 100 units
    • Breadth: 50 ± 0.25 units
  • Calibration Marks:
    • Length: 7.5 units on the longer axis
    • Spacing: 85.4 ± 0.25 units between marks
  • Ruling Lines:
    • Thickness ≤ 1/5 unit (max 1/500th grid length)
    • Uniformity within ±20%
  • Circle Diameters (units ± tolerance):
Circle No.Diameter (units)Tolerance
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14

Notes:

  • Lengths measured between centers of ruling lines.
  • Circle diameters measured between outside edges.
  • Grid subdivided as per figure with specified area ratios.
  • Pattern includes rectangular grid, calibration marks, open and solid circles (numbered 1–7), grid length label, and manufacturer's details.

Unit Length Calculation:

[ \text{Unit length} = \frac{\text{Grid length}}{100} ]


graph TB
    A[Rectangular Grid] --> B[Calibration Marks (7.5 units)]
    A --> C[Subdivision per Figure]
    A --> D[Open Circles (7 sizes)]
    A --> E[Solid Circles (7 sizes)]
    B --> F[Spacing Between Marks: 85.4 ± 0.25 units]
    D --> G[Circle Diameters with Tolerances]

This framework ensures accurate and reproducible graticule design following IS 5257.

2.2Arrangement of Circles and Ruling Lines

IS 5257: Layout of Circles and Rulings — Key Features

1. Graticule Pattern (Clause 2.2)

  • Rectangular grid subdivided as per figure.
  • Two calibration marks on the longer axis, crossing at intersections extended by two units.
  • Seven open and seven solid circles, numbered 1 through 7.
  • Label "GRID LENGTH" with actual measurement in millimeters.
  • Manufacturer's name and standard number included.

2. Construction Details (Clause 2.3.1)

  • Boundary lines extend by 2.5 units beyond the grid edges to intersect.
  • Calibration marks length: 7.5 units, symmetrically located.
  • Subdivision lines create defined areas as per specification.
  • Ruling line thickness ≤ 1/5 unit (1/500th of grid length), with ±20% tolerance.

3. Dimensions (Table 1)

ComponentDimension (units)Tolerance (units)
Grid length100-
Grid breadth50±0.25
Calibration marks spacing85.4±0.25
Circle Diameters:
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14
  • Unit length equals 1/100th of grid length.
  • Tolerances apply to both eyepiece and screen graticules, with minimum precision of 5 microns.

Unit Length Formula:

[ \text{Unit length} = \frac{\text{Grid length (mm)}}{100} ]


Conceptual Visualization:

graph LR
    A[Rectangular Grid] --> B[Calibration Marks]
    B --> C[Subdivisions]
    B --> D[Open and Solid Circles]
    D --> E[Circle Diameter Tolerances]

This layout supports precise, standardized graticule production.

2.3Proportional Dimensions of Grid and Circles

IS 5257: Proportional Dimensions for Grid and Circles

Specifications (Clause 2.3 & Table 1)

  • Base unit: Defined as 1/100th of the grid length.
  • Grid length: Fixed at 100 units.
  • Grid breadth: 50 ± 0.25 units.
  • Calibration marks interval: 85.4 ± 0.25 units.
  • Circle diameters (units ± tolerance):
Circle No.Diameter (units)Tolerance
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14

Additional Construction Notes (Clause 2.3.1)

  • Boundary lines extend 2.5 units beyond the grid to intersect.
  • Calibration marks on the longer axis are 7.5 units long and symmetric.
  • Ruling lines have thickness ≤ 1/5 unit, consistent within ±20%.
  • Lengths measured between centers of ruling lines; circle diameters between outer edges.

Actual Grid Dimensions (Clause 2.4)

  • Grid length must be within ±2% of the nominal value.
  • Screen graticule grid length minimum: 106.1 mm.

Diagrammatic Representation:

graph TD
    A[Rectangular Grid] -->|Length = 100 units| B[Grid Length]
    A -->|Breadth = 50 ± 0.25 units| C[Grid Breadth]
    A --> D[Calibration Marks]
    D -->|Length = 7.5 units| E[Long Axis Marks]
    A --> F[Open Circles]
    F --> G1[Circle 1: 1.41 ± 0.03 units]
    F --> G2[Circle 2: 2.00 ± 0.04 units]

This ensures accurate, standardized dimensions for graticules.

2.4Grid Dimension Specifications

IS 5257 - Grid Dimensions for Graticules

Essential Specifications & Formulas:

  • Grid length (long side): 100 units (with base unit = 1/100th of grid length).
  • Grid breadth (short side): 50 ± 0.25 units.
  • Calibration marks spacing: 85.4 ± 0.25 units.
  • Line width of rulings: ≤ 1/5 unit (max 1/500th of grid length), uniform within ±20%.
  • Calibration mark length: 7.5 units on the longer axis, symmetrically placed.
  • Boundary lines: Extended by 2.5 units beyond the grid to intersect.

Circle Diameters (units ± tolerance):

Circle No.Diameter (units)Tolerance
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14

Additional Information:

  • Actual grid length must be within ±2% of the marked length.
  • Minimum grid length for screen graticules: 106.1 mm.
  • Measurements are taken center to center for ruling lines, and outside edge to outside edge for circle diameters.

Unit Length Formula:

[ \text{Unit length} = \frac{\text{Grid length}}{100} ]


graph LR
A[Grid Length = 100 units]
B[Grid Breadth = 50 ± 0.25 units]
C[Calibration Marks Length = 7.5 units]
D[Calibration Marks Distance = 85.4 ± 0.25 units]
E[Circle Diameters per Table]
A --> B
A --> C
A --> D
A --> E

This guarantees precise and consistent graticule dimensions for accurate measurement.

3Marking Guidelines for Graticules

IS 5257 - Marking Protocols and Dimension Specifications

1. Marking Requirements (Clause 3.1)

  • Manufacturer’s name, trademark, or identifying mark must be present.
  • The Indian Standard number (IS 5257) should be clearly indicated.
  • Optional inclusion of ISI Certification Mark, following ISI regulations, to ensure quality assurance.

2. Graticule Dimension and Features (Clause 2.3.1 & Table 1)

ComponentDimension (units)Allowable Deviation
Grid length100-
Grid breadth50±0.25
Calibration mark spacing85.4±0.25
Circle diameters:
- Circle 11.41±0.03
- Circle 22.00±0.04
- Circle 32.83±0.06
- Circle 44.00±0.08
- Circle 55.66±0.10
- Circle 68.00±0.12
- Circle 711.31±0.14
  • Ruling line thickness ≤ 1/5 unit (1/500th of grid length), uniform within ±20%.
  • Calibration marks length: 7.5 units, symmetrically placed on the longer axis.
  • Boundary lines extend 2.5 units beyond intersections.

3. Pattern Components (Clause 2.2)

  • Rectangular grid subdivided as per figure.
  • Two calibration marks crossing the longer axis.
  • Sets of seven open and seven solid circles numbered 1 to 7.
  • Markings include "GRID LENGTH" with actual length in mm.
  • Manufacturer’s name and IS number.

Visual Summary (Mermaid Diagram)

graph TD
    A[Rectangular Grid] --> B[Boundary lines extended]
    B --> C[Calibration marks]
    C --> D[Circles]
    D --> E[Markings and Labels]

This marking protocol ensures traceability and quality, aligning with IS 5257 standards.

3.1Manufacturer Identification and Standard Labeling

IS 5257 Marking Requirements for Manufacturer and Standard Number (Clause 3.1):

  • The graticule must display the manufacturer’s name, trademark, or other identification.

  • The Indian Standard code (IS 5257) must be visibly marked.

  • ISI Certification Mark (Optional):

    • Signifies adherence to IS standards.
    • Indicates inspection, testing, and quality control compliance.
    • Usage governed by rules of the Indian Standards Institution.

Reference Table: Key Dimensions of Standard Graticule (Clause 2.3.1)

FeatureValue ± Tolerance (units*)
Grid length100
Grid breadth50 ± 0.25
Calibration mark spacing85.4 ± 0.25
Circle diameters:
- Circle 11.41 ± 0.03
- Circle 22.00 ± 0.04
- Circle 32.83 ± 0.06
- Circle 44.00 ± 0.08
- Circle 55.66 ± 0.10
- Circle 68.00 ± 0.12
- Circle 711.31 ± 0.14

*Unit based on grid length (1 unit = 1/500th of grid length).


Schematic Diagram:

graph TD
  A[Rectangular Grid] --> B[Length = 100 units]
  A --> C[Breadth = 50 ± 0.25 units]
  A --> D[Calibration Marks (7.5 units)]
  A --> E[Circles of Various Diameters]
  E --> F[Diameters Range 1.41 to 11.31 units ± tolerances]
  A --> G[Ruling Line Width ≤ 1/5 unit]

Manufacturer and IS number marking provide traceability and guarantee compliance with IS 5257, critical for accurate particle size measurement.

4Related Standards and References

IS 5257: Relevant Standards and References

Associated Indian Standards for Sieving and Particle Size Analysis:

IS NumberTitlePrice (Rs)
IS 460-1962Test sieves (revised)8.50
IS 1568-1960Wire gauge for general applications1.50
IS 1607-1960Procedures for dry sieving5.50
IS 2405-1963Wire cloth and perforated plates for industrial sieves2.50
IS 3150-1966Hexagonal wire mesh for general use (revised)1.50
IS 4124-1967Glossary of powder-related terminology3.50
IS 4879-1968Method for subdividing gross powder samples for particle size analysis5.50
IS 4961-1968Particle size determination by air elutriation methods7.50

Summary of Graticule Specifications (IS 5257):

  • Grid length: 100 units
  • Grid breadth: 50 ± 0.25 units
  • Calibration mark spacing: 85.4 ± 0.25 units
  • Circle diameters:
    • 1: 1.41 ± 0.03 units
    • 2: 2.00 ± 0.04 units
    • 3: 2.83 ± 0.06 units
    • 4: 4.00 ± 0.08 units
    • 5: 5.66 ± 0.10 units
    • 6: 8.00 ± 0.12 units
    • 7: 11.31 ± 0.14 units
  • Ruling line thickness: ≤ 1/5 unit (max 1/500th grid length), consistent within ±20%

Marking Requirements (Clause 3.1):

  • Manufacturer’s name or trademark
  • IS Standard number (IS 5257)

Purpose:

These standards ensure uniformity and accuracy in particle size determination for powders.

Annex AIllustration of Indian Standard Graticule (Figure 1)

IS 5257 - Standard Graticule Illustration (Figure 1) Key Details

1. Pattern and Components (Clause 2.2)

  • Rectangular grid with length 100 units and breadth 50 ± 0.25 units.
  • Two calibration marks on the longer axis, each 7.5 units in length.
  • Seven open circles and seven solid circles, labeled 1 through 7.
  • Markings include "GRID LENGTH" followed by actual length in millimeters, manufacturer’s name, and standard number.

2. Construction Features (Clause 2.3.1)

  • Grid boundary lines extend 2.5 units beyond grid to form intersections.
  • Distance between calibration marks: 85.4 ± 0.25 units.
  • Ruling line thickness ≤ 1/5 unit (maximum 1/500th of grid length), consistent within ±20%.
  • Circle diameters measured between outer edges of ruling lines.

3. Circle Diameters (Table 1)

Circle No.Diameter (units)Tolerance (±)
11.410.03
22.000.04
32.830.06
44.000.08
55.660.10
68.000.12
711.310.14

4. Notes:

  • Lengths and distances measured center-to-center of ruling lines.
  • Circle diameters measured outside edge to outside edge.
  • Graticule printed with black rulings on a diffusing background for screen graticules.

flowchart LR
    A[Rectangular Grid]
    A --> B[Length = 100 units]
    A --> C[Breadth = 50 ± 0.25 units]
    A --> D[Calibration Marks (7.5 units)]
    A --> E[Seven Open and Seven Solid Circles]
    E --> F[Diameters as Specified in Table 1]

This figure represents the official design for graticules as per IS 5257.

Popular Questions About IS 5257

?What are the exact dimensional tolerances for the grid and circles in the graticule?

The dimensional tolerances specified in IS 5257 for the graticule are as follows:

  • Base unit is 1/100th of the rectangular grid length.
  • Grid length is fixed at 100 units (reference).
  • Grid breadth is 50 ± 0.25 units.
  • Distance between calibration marks is 85.4 ± 0.25 units.
  • Circle diameters with their tolerances (units ±):
Circle No.DiameterTolerance
11.41±0.03
22.00±0.04
32.83±0.06
44.00±0.08
55.66±0.10
68.00±0.12
711.31±0.14
  • Ruling line thickness must be ≤ 1/5 unit (i.e., ≤ 1/500th of grid length) and consistent within ±20%.
  • Actual grid length tolerance is ±2% of the marked value; minimum grid length for screen graticules is 106.1 mm.
  • Length measurements are between centers of ruling lines, while circle diameters are measured between outside edges.

This precision ensures accurate scaling and reliable calibration for microscopy-based particle size measurements.

?How should the graticules be marked to comply with IS 5257?

To meet IS 5257 marking requirements, graticules must include:

  • The manufacturer's name, trademark, or other identifying mark.
  • The Indian Standard number (IS 5257) clearly displayed.
  • Optionally, the ISI Certification Mark may be used, subject to licensing and adherence to Indian Standards Institution regulations.

Graticule features to be marked include:

ComponentDimension (units)Tolerance
Grid length100±2%
Grid breadth50±0.25
Calibration marks length7.5-
Distance between marks85.4±0.25
Circle diameters (1 to 7)See specificationsSee tolerances
Ruling line width≤ 1/5 unit±20% variation

The markings must be legible and permanent, ensuring traceability and quality conformity according to IS 5257.

?What materials and construction methods are specified for eyepiece and screen graticules?

IS 5257 specifies that eyepiece and screen graticules must be constructed with:

  • Black ruling lines on a transparent substrate for eyepiece graticules.
  • For screen graticules, the background should be a diffusing screen to enhance visibility.

Construction specifics include:

  • Ruling line thickness ≤ 1/5 unit (1/500th of grid length), uniform within ±20%.
  • Boundary lines extended by 2.5 units beyond the grid to intersect.
  • Calibration marks on the longer axis with lengths of 7.5 units, symmetrically positioned.
  • Grid dimensions: length of 100 units and breadth of 50 ± 0.25 units.
  • Distance between calibration marks: 85.4 ± 0.25 units.
  • Circle diameters as per defined specifications with tolerances.

Measurements are made center to center for ruling lines and outside edges for circles, ensuring precise and reproducible particle size measurement under the microscope.

?How does IS 5257 ensure compatibility with particle size determination for powders passing a 75-micron sieve?

IS 5257 ensures compatibility with particle size analysis of powders passing through a 75-micron sieve by:

  • Defining eyepiece and screen graticules designed specifically for optical microscopes to measure particles below 75 microns.
  • Requiring particle samples to be prepared on slides and viewed under transmitted light microscopy.
  • Mandating comparison of particle images against calibrated reference circles on the graticule to determine size accurately.
  • Aligning the standard with international references such as BS 3625:1963 for consistency.

This facilitates a reliable bridge between sieve analysis and microscopic sizing, providing accurate particle size distribution data for fine powders.

?What calibration marks are required on the graticule for accurate particle size measurement?

According to IS 5257, the calibration marks on the graticule must have the following features for precise particle size measurement:

  • Two calibration marks drawn across the longer axis of the rectangular grid.
  • Each calibration mark must be 7.5 units in length and symmetrically placed.
  • The spacing between the two calibration marks is 85.4 ± 0.25 units, measured center to center.
  • The grid has a length of 100 units and a breadth of 50 ± 0.25 units.
  • Boundary lines of the grid extend by 2.5 units to intersect.
  • Ruling line thickness should be consistent within ±20% and must not exceed 1/5 unit (1/500th of grid length).

This configuration ensures accurate scaling and measurement of particle sizes via optical microscopy in line with IS 5257.